|
|
|
National Stock Number: 6530-01-431-5146
Federal Supply Class: 6530
National Item Identification Number: 014315146
Description: VENTILATOR,VOLUME,PORTABLE
Detail: AN ITEM DESIGNED TO PROVIDE VENTILATION TO A PATIENT IN TRANSPORT AND/OR DURING INSTITUTIONAL CARE. IT MAY BE IN A TRANSPORT CASE AND PROVIDE TIME CYCLED, PRESSURE LIMITED, PISTON DRIVEN CONSTANT FLOW, BE MICROPROCESSOR CONTROLLED, WITH THREE MODES, CONTROL, ASSIST-CONTROL, SYNCHRONOUS INTERMITTENT MANDATORY VENTILATION PROVIDING SENSITIVITY CONTROL, RECHARGEABLE BATTERY, HOSES, TUBING, CONNECTORS, PATIENT VALVE, AND OTHER RELATED FEATURES.
|
Manufacturer Information:
| 96-0337 | 89875 | DEFENSE LOGISTICS AGENCY DIRECTORATE OF MEDICAL MATERIEL DEFENSE SUPPLY CENTER PHILADELPHIA |
|
Techincal Specification:
| CURRENT TYPE | AC OR DC | | FURNISHED ITEMS AND QUANTITY | OPERATORS, 2; SERVICE MANUALS, 2 | | OVERALL HEIGHT | 21.000 INCHES NOMINAL | | OVERALL LENGTH | 18.000 INCHES NOMINAL | | OVERALL WIDTH | 18.000 INCHES NOMINAL | | PART NAME ASSIGNED BY CONTROLLING AGENCY | VENTILATOR VOLUME PTBL: | | PORTABILITY | PORTABLE | | SPECIAL TEST FEATURES | JET AIRCRAFT RANDOM VIBRATON TESTING | | SUPPLEMENTARY FEATURES | UNIT SHALL BE COMPATIBLE WITH THE FOLLOWING REGULATORS: PRESSURE REGULATOR, OXYGEN, FOR D-CYLINDERS, NSN 6680-01-505-1310 AND PRESSURE REGULATOR, OXYGEN, FOR H-CYLINDERS, NSN 6680-01-505-1303 | | SPECIAL FEATURES | THE UNIT IS A TRANSPORT VENTILATOR FOR AIR AND GROUND MOVEMENT OF PATIENTS. IT IS A SELF-CONTAINED SYSTEM CAPABLE OF SUPPLYING 21% TO 100% OXYGEN WHEN PROVIDED 50 PSI OXYGEN SOURCE. THE BREATHING CIRCUIT USES STANDARD DISPOSABLE UNIVERSAL 22 MM TUBING SET. FUNCTION MODES: ASSIST/CONTROL; INTERMITTENT MANDATORY VENTILATION (IMV); CONTINUOUS POSITIVE AIRWAY PRESSURE (CPAP); AND POSITIVE END EXPIRATORY PRESSURE (PEEP). ALARMS: APNEA, HIGH PRESSURE, LOW PRESSURE (DISCONNECT), EXTERNAL POWER (LOW/FAIL), BATTERY (LOW/FAIL), LOW GAS SOURCE PRESSURE, LOW PEEP AND LOW EXHALE VOLUME. VISUAL: BREATHS PER MIN., PEAK INSPIRATORY PRESSURE (PIP), MEAN AIRWAY PRESSURE (MAP), PEEP, EXHALE VOLUME, SPONTANEOUS PATIENT RATE, AND ALARM STATUS. THE UNIT ALLOWS VENTILATION OF BOTH CHILDREN (1-12 YEARS) AND ADULTS. ELECTRICAL: 115/230 VAC, 50/60/400 HZ, 11-30 VDC; JOINT AIR CERTIFICATION FOR FIXED & ROTARY WING AIRCRAFT FOR COMBINED ESSECTIAL CHARACTERISTICS | | TEST DATA DOCUMENT | 96906-1472 STANDARD (INCLUDES INDUSTRY OR ASSOCIATION STANDARDS, INDIVIDUAL MANUFACTUREER STANDARDS, ETC.). AND 80204-ES1-1999 STANDARD (INCLUDES INDUSTRY OR ASSOCIATION STANDARDS, INDIVIDUAL MANUFACTUREER STANDARDS, ETC.). |
|
|